Measurements and Results
R&S
®
FSW
352
User Manual 1173.9411.02 ─ 43
Demodulation begins immediately with the next measurement. During the initial
measurement, demodulation is performed for the entire measurement span; during
the final measurement only the detected peak marker positions are demodulated
(for the defined dwell time).
12. Increase the number of sweep points for the EMI measurement.
a) Press the [SWEEP] key on the front panel.
b) Select the "Sweep Config" softkey.
c) Set the number of "Sweep Points" so that the distance between two sweep
points is smaller than RBW/3.
13. Optionally, select or configure limit lines against which the marker results are
checked.
a) Press the [Lines] key and then the "Lines Config" softkey, then select the "Lines
Config" tab.
b) In the "Line Config" dialog box, select the "View Filter" option: "Show Compati-
ble".
All stored limit lines with the file extension
.LIN
in the
limits
subfolder of the
main installation folder of the instrument that are compatible to the current EMI
measurement settings are displayed in the overview.
c) Select the "Check Traces" setting for a limit line in the overview and select the
trace numbers to be included in the limit check. One limit line can be assigned
to several traces.
14. Define a suitable unit for the measured values, as the default unit dBm is not suita-
ble for EMI measurements, or select a transducer.
To change the unit:
Press the [AMPT] key, then select the "Amplitude Config" softkey and, in the
"Amplitude" dialog box, select the required unit.
To select a transducer:
a) Press the [SETUP] key.
b) Select the "Transducer" softkey.
c) In the "Transducer" dialog box, set the "View Filter" to "Show Compatible" to
determine the available transducers for the current EMI measurement setup.
d) Select a transducer line in the overview and select the "Active" setting for it.
15. Press the [RUN SINGLE] key to start a new EMI measurement.
If activated, a peak search is performed. For each active marker, a final measure-
ment is performed using the specified detector for the specified dwell time. If acti-
vated, the signal is demodulated at the active marker positions.
The specified traces to be checked are compared with the active limit lines. The
status of the limit check for the final measurement is indicated in the Result Sum-
mary.
Electromagnetic Interference (EMI) Measurement