Measurements and Results
R&S
®
FSW
332
User Manual 1173.9411.02 ─ 43
7.13.1
About the EMI Measurement
EMI measurements can be very time-consuming, especially if weighting detectors are
required for the measurement. In addition, EMC testing often requires various proce-
dures to locate local EMI maxima. Such procedures are, for example, movements of
an absorbing clamp, variations in the height of the test antenna or the rotation of the
DUT.
Covering all test setups with one of the (slow) EMI weighting detectors over the
required frequency range would lead to very high measurement times.
Splitting the measurement procedure into several stages, however, can eliminate this
problem.
The first stage, or peak search, is used to get a rough idea about the location of peak
levels that may indicate interference over the required frequency range. This stage
uses a detector that allows for a fast sweep time, e.g. the peak detector.
During the second stage, or final test, the R&S
FSW performs the actual EMC test, a
refined measurement with detectors designed for and required by EMC standards. To
keep measurement times brief, the R&S
FSW performs a final measurement only on
frequencies you have marked with a marker or delta marker. You can assign a different
detector to every marker and thus test a particular frequency easily for compliance.
Optionally, you can activate continuous demodulation of the signal during the initial
measurement and at the peak marker positions during the final test.
After the final measurement, you can check the signal levels against specified limits.
7.13.2
EMI Measurement Results
As the result of an R&S
FSW EMI measurement, the measured signal levels and
active markers are displayed in a Spectrum diagram.
Figure 7-57: EMI measurement results
Electromagnetic Interference (EMI) Measurement
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...