Common Analysis and Display Functions
R&S
®
FSW
613
User Manual 1173.9411.02 ─ 43
To store data permanently, select an external storage location such as a USB memory
device.
For details, see
Chapter 5.1.7, "Protecting Data Using the Secure User Mode"
Remote command:
File Type ← Export Trace to ASCII File
Determines the format of the ASCII file to be imported or exported.
Depending on the external program in which the data file was created or is evaluated,
a comma-separated list (
CSV
) or a plain data format (
DAT
) file is required.
Remote command:
Decimal Separator ← Export Trace to ASCII File
Defines the decimal separator for floating-point numerals for the data export/import
files. Evaluation programs require different separators in different languages.
Remote command:
File Explorer ← Export Trace to ASCII File
Opens the Microsoft Windows File Explorer.
Remote command:
not supported
Importing Traces
Trace data that was stored during a previous measurement can be imported to the
Spectrum application, for example as a reference trace.
Import All Traces/Import to Trace ← Importing Traces
If the import file contains more than one trace, you can import several traces at once,
overwriting the existing trace data for any active trace in the result display with the
same trace number. Data from the import file for currently not active traces is not
imported.
Alternatively, you can import a single trace only, which is displayed for the trace num-
ber specified in "Import to Trace". This list contains all currently active traces in the
result display. If a trace with the specified number exists in the import file, that trace is
imported. Otherwise, the first trace in the file is imported (indicated by a message in
the status bar).
Remote command:
Import ASCII File to Trace ← Importing Traces
Loads one trace or all traces from the selected file in the selected ASCII format (
.DAT
or
.CSV
) to the currently active result display.
Importing and Exporting Measurement Results for Evaluation
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...