Remote Commands
R&S
®
FSW
1146
User Manual 1173.9411.02 ─ 43
Reading out data in binary format is quicker than in ASCII format. Thus, binary format
is recommended for large amounts of data.
14.8.2.6
Importing and Exporting Traces
............................................................................................1146
............................................................................................1146
............................................................................................ 1147
.........................................................................................1147
....................................................................................... 1147
FORMat:DEXPort:FORMat
<FileFormat>
Determines the format of the ASCII file to be imported or exported. Depending on the
external program in which the data file was created or will be evaluated, a comma-sep-
arated list (CSV) or a plain data format (DAT) file may be required.
Parameters:
<FileFormat>
CSV | DAT
*RST:
DAT
Example:
FORM:DEXP:FORM CSV
Manual operation:
See
FORMat:DEXPort:TRACes
<Selection>
This command selects the data to be included in a data export file (see
For details on exporting data see
Chapter 9.6.2, "Trace/Data Ex/Import"
Parameters:
<Selection>
SINGle | ALL
SINGle
Only a single trace is selected for export, namely the one speci-
fied by the
command.
ALL
Selects all active traces and result tables (e.g. Result Summary,
marker peak list etc.) in the current application for export to an
ASCII file.
The <trace> parameter for the
command is ignored.
*RST:
SINGle
Manual operation:
See
"Export all Traces and all Table Results"
Analyzing Measurements (Basics)