Remote Commands
R&S
®
FSW
1275
User Manual 1173.9411.02 ─ 43
<fli>
1..n
Index in frequency response file list
Use
[SENSe:]CORRection:FRESponse<si>:BASeband:
to determine the maximum index
number.
Example:
SENS:CORR:FRES:USER:FLIS2:DATA:PHAS?
Usage:
Query only
Manual operation:
See
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:FLISt<fli>:INSert
<FilePath>
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:FLISt<fli>:INSert
<FilePath>
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:INSert
<FilePath>
Loads a frequency response (
.fres
) file to the current configuration. The maximum
number of files per configuration is 15. The new file is added below the entry specified
by the
<fli>
index. All other entries with a higher suffix are moved down by one posi-
tion.
Suffix:
<si>
.
1..n
irrelevant
<fli>
1..n
Index in frequency response file list
Use
[SENSe:]CORRection:FRESponse<si>:BASeband:
to determine the maximum index
number.
Parameters:
<FilePath>
string
Path and file name
The default directory for
.fres
files is C:\R_S\INSTR\USER
\Fresponse.
Example:
SENS:CORR:FRES:USER:FLIS2:INS 'C:\FRes.fres'
Manual operation:
See
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:FLISt<fli>:MAGNitude[:
STATe]
<State>
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:FLISt<fli>:MAGNitude[:
STATe]
<State>
[SENSe:]CORRection:FRESponse<si>:USER:FLISt<fli>:MAGNitude[:STATe]
<State>
Activates or deactivates the use of the correction data in the selected file for magnitude
results.
Configuring the R&S
FSW