Measurements and Results
R&S
®
FSW
229
User Manual 1173.9411.02 ─ 43
Connect/Disconnect ← Signal Generator IP Address
The R&S
FSW attempts to establish a connection to the signal generator, or discon-
nects it.
If an instrument is connected, the following information is displayed:
●
Device type
●
Name and serial number
●
Connection state
Remote command:
CONFigure:GENerator:CONNection[:STATe]
CONFigure:GENerator:CONNection:CSTate?
Generator Control State (FSW -> SMW)
Activates or disables control of the signal generator by the R&S
FSW.
If a connection was defined in another application, the connection is maintained when
you switch to the Spectrum application. However, generator control is disabled to pro-
tect the DUT from possibly erroneous or damaging settings. Check the settings, then
enable the control state.
Note:
While generator control is active, you cannot change the connection information.
When you switch applications, the control is disabled in the other applications. Only
one application can control a generator at any time.
As long as the R&S
FSW controls the signal generator, any (general) generator set-
tings you define on the R&S
FSW are automatically adapted on the generator. The
opposite does not apply, that is: changes you make on the generator are not automati-
cally applied to the R&S
FSW. The green LED for a previous successful transmission
Query all Generator Setup Settings from Generator
to apply the generator settings on the R&S
FSW.
Remote command:
CONFigure:GENerator:NPRatio:CONTrol[:STATe]
Path RF/ Path BB
Selects the RF signal path and indicates the BB signal path of the generator to be used
for signal generation.
Remote command:
CONFigure:GENerator:TARGet:PATH:RF
CONFigure:GENerator:TARGet:PATH:BB?
RF Output State
To protect the instrument from possibly erroneous or damaging settings, you must
manually activate the RF output on the signal generator to start providing a signal.
Check all settings on the signal generator, in particular the level settings, before acti-
vating the RF output.
A red LED on the "Generator Control" tab indicates a setting error on the generator.
Remote command:
CONFigure:GENerator:RFOutput[:STATe]
CONFigure:GENerator:NPRatio:RFOutput:STATe:CSTate?
Noise Power Ratio (NPR) Measurement