Common Measurement Settings
R&S
®
FSW
465
User Manual 1173.9411.02 ─ 43
the instrument, occupying a large amount of memory, and each sweep point increases
the overall measurement time.
The number of sweeps to be performed in single sweep mode is defined by the
"Sweep Count". Values from 0 to 200000 are allowed. If the values 0 or 1 are set, one
sweep is performed. The sweep count is applied to all the traces in a diagram.
If the trace configurations "Average", "Max Hold" or "Min Hold" are set, the "Sweep/
Average Count" also determines the number of averaging or maximum search proce-
dures (see
"Analyzing Several Traces - Trace Mode"
For details on how the number of sweep points and the sweep count affect the trace
results on the screen, see
"Mapping Samples to sweep Points with the Trace Detector"
8.5.1.9
How Often Data is Measured: Sweep Mode
How often the spectrum is swept depends on the sweep mode. Either a certain number
of sweeps can be defined ("Sweep Count") which are performed in "Single Sweep"
mode, or the sweep is repeated continuously ("Continuous Sweep" mode).
By default, the data is collected for the specified number of sweeps and the corre-
sponding trace is displayed. When the next sweep is started, the previous trace is
deleted.
However, the data from a single sweep run can also be retained and displayed
together with the new data ("Continue Single Sweep" mode). This is particularly of
interest when using the trace configurations "Average" or "Max Hold" to take previously
recorded measurements into account for averaging/maximum search (see
8.5.2
Bandwidth, Filter and Sweep Settings
Access
: "Overview" > "Bandwidth"
The remote commands required to define these settings are described in
ter 14.7.2, "Configuring Bandwidth and Sweep Settings"
How to perform a basic sweep measurement is described in
Perform a Basic Sweep Measurement"
Bandwidth, Filter and Sweep Configuration