Publication No. 980935 Rev. A
3152B User Manual
EADS North America Test and Services
Programming Reference 5-59
Table 5-7, Apply Control Commands
Keyword
Parameter Form
Default
3152A
HS
[:SOURce]
:APPLy
FREQ,AMPL,OFFS
9
0300C
:SINusoid
FREQ,AMPL,OFFS,PHAS,POW
9
0301E
:TRIangle
FREQ,AMPL,OFFS,PHAS,POW
9
0302E
:SQUare
FREQ,AMPL,OFFS,DCY
9
0303D
:PULSe
FREQ,AMPL,OFFS,DEL,WID,LEE,TRE
9
0304G
:RAMP
FREQ,AMPL,OFFS,DEL,LEE,TRE
9
0305F
:SINC
FREQ,AMPL,OFFS,CYC
9
0310D
:GAUssian
FREQ,AMPL,OFFS,EXP
9
0306D
:EXPonential
FREQ,AMPL,OFFS,EXP
9
0307D
:DC
DC_AMPL
9
0308A
:USER
SEG<n>,SCLK,AMPL,OFFS
9
0309D
APPLy <freq>,<ampl>,<offs>(?)
HS Syntax: 0300C
<float>,<float,<float>(?)
Description
This command changes the waveform function to standard and programs the frequency, amplitude, and offset
for the selected standard waveform. This command affects the output regardless of the current output
function. For example, if the 3152B is in FM mode, the 3152B will stop generating FM and will revert to the
currently selected standard waveform with values of frequency, amplitude and offset as specified by this
command.
Parameters
Name
Range
Type
Default Description
<freq>
10e-3 to
100e6
Numeric
1e6
Programs the frequency of the standard waveform in
units of Hz.
<ampl>
10e-3 to
16
Numeric
5
Programs the amplitude of the standard waveform in
units of Volts.
<offs>
-7.992 to
7.992
Numeric
0
Programs the offset of the standard waveform in units of
Volts.
Response
The 3152B returns the current frequency, amplitude and offset setting as in the following example: 1e6,5,0.
APPLy:SINusoid <freq>,<ampl>,<offs>,<phas>,<power>(?)
HS Syntax: 0301E
<float>,<float,<float>,<float>,<integer>(?)
Description
This command changes the waveform function to standard sine and programs the frequency, amplitude,
offset, start phase, and power exponent simultaneously. This command affects the output regardless of the
currently selected operating mode. For example, if the 3152B is in FM mode, the 3152B will stop generating
FM and will revert to the currently selected standard waveform with parametric values as specified by this
command.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com