Publication No. 980935 Rev. A
3152B User Manual
EADS North America Test and Services
Programming Reference 5-41
TRANsition
Discrete
Selects a transition pulse with a programmable
threshold and slope to first turn the gate on and then to
turn it off.
Response
The 3152B returns LEV or TRAN depending on the selected option.
TRIGger:GATE {OFF|ON|0|1}(?)
HS Syntax: 1150A<0|1>(?)
Description
Toggle gated run mode on or off. Affect the 3152B only after it is set to INIT:CONT OFF mode.
Parameters
Range
Type
Default
Description
0-1
Discrete
0
Turns gate run mode off or on.
Response
The 3152B returns 0 or 1 depending on the selected option.
TRIGger:LEVel<level>(?)
HS Syntax: 1180A<float>(?)
Description
The trigger level command sets the threshold level at the trigger input connector. The trigger level command
affects the generator when it is an in interrupted run mode which can be selected using the INIT:CONT OFF
command.
Parameters
Name
Range
Type
Default
Description
<level>
-10 to 10
Numeric
0
Programs the trigger threshold. The value affects the
front panel input only.
Response
The 3152B returns the current trigger threshold value.
TRIGger:SOURce:ADVance {EXTernal|INTernal|TTLTrg<n>|ECLT1|BUS}(?)
HS Syntax: 1140A<4|3|8|9|10|11|12|13|14|15|16|17>(?)
Description
This selects the trigger source forthe 3152B. The source advance command affects the generator only after it
has been programmed to operate in an interrupted run mode which can be selected using the INIT:CONT
OFF command.
Parameters
Name
Type
Default
Description
EXTernal
Discrete
EXT
Selects the front panel TRIG IN input as the trigger
source.
INTernal
Discrete
Activates the built-in internal trigger generator. BUS and
external triggers are ignored. The period of the internal