3152B User Manual
Publication No. 980935 Rev. A
Table 6-20, Sine Wave Flatness Test, DDS Output – at 10 V
3152B Sine
Frequency
Error Limits
Oscilloscope Reading
Pass
Fail
1 kHz
6 Divisions
Reference
X
X
1 MHz
6 ±0.3 Divisions
10 MHz
6 ±0.3 Divisions
50 MHz
6 ±0.9 Divisions
Trigger Operational
Characteristics
This tests the operation of the trigger circuit including tests for the
triggered, gated, and counted burst run modes. It also tests the
operation of the triggered advance options, the delayed trigger, the re-
trigger functions, as well as the trigger input level and slope sensitivity.
Trigger, Gate, and
Burst
Characteristics
Equipment: Oscilloscope, function generator, counter
Preparation:
1. Configure the Oscilloscope as follows:
Termination: 50
Ω
, 20 dB feedthrough attenuator at the
oscilloscope input
Setup:
As required for the test
2. Configure the counter as follows:
Function:
TOT B
Trigger Level: 100 mV
3. Connect 3152B output to the oscilloscope input
4. Configure the function generator as follows:
Frequency 1
MHz
Run Mode:
As required by the test
Wave:
2 V Square
5. Connect the function generator output to the 3152B TRIG IN
connector
6. Configure the 3152B as follows:
Frequency: 25
MHz
Waveform: Sine
wave
Burst Count:
1e6 counts
Amplitude: 1
V
Trigger Source: External
Output: On
Test Procedure
1. Perform trigger and gate tests using Table 6-21.
Table 6-21, Trigger, Gate, and Burst Characteristics
3152B Run
External Trigger
Oscilloscope/Counter
Mode Pulse
Reading
Pass
Fail
Triggered
1 MHz, Continuous
Triggered waveform
Gated – Transition
1 MHz, Continuous
Gated by transition Waveform
Gated – Level
1 MHz, Continuous
Gated by level Waveform
Burst
Single shot
Burst, 1e6 waveforms
6-12 Performance Checks
EADS North America Test and Services