Publication No. 980935 Rev. A
3152B User Manual
EADS North America Test and Services
Programming Reference 5-95
DPULse:DOUBle:DELay <d_delay>(?)
HS Syntax: 0145A<float>(?)
Description
Programs the delay between two adjacent pulses when double pulse mode is selected. Otherwise, the double
pulse delay has no effect on the pulse structure.
Parameters
Name
Range
Type
Default
Description
<d_delay>
0 to 1e3
Numeric
1e-3
Sets the delay between two adjacent pulses for the
double pulse mode in units of seconds. Note that the
sum of all parameters, including the pulse delay time
must not exceed the programmed pulse period and
therefore, it is recommended that the pulse period be
programmed before all other pulse parameters.
Response
The 3152B returns the current double pulse delay value in units of seconds.
DPULse:LEVel:HIGH <high>(?)
HS Syntax: 0147A<float>(?)
Description
Programs the high level for the pulse shape. Note that the same level is retained for the second pulse in the
double pulse mode.
Parameters
Name
Range
Type
Default
Description
<high>
-7.992 to 8
Numeric
5
Sets the pulse high level in units of Volts. Note that the
high level setting must be higher than the low level
setting. Also note that high to low level value must be
equal or larger than 8 mV.
Response
The 3152B returns the current low level value in unit of Volts.
DPULse:LEVel:LOW <low>(?)
HS Syntax: 0148A<float>(?)
Description
Programs the low level for the pulse shape. Note that the same level is retained for the second pulse in the
double pulse mode.
Parameters
Name
Range
Type
Default
Description
<low>
-8 to 7.992
Numeric
0
Sets the pulse low level in units of Volts. Note that the
low level setting must be smaller than the high level
setting. Also note that low to high level value must be
equal or larger than 8 mV.
Response
The 3152B returns the current high level value in unit of Volts.