Publication No. 980935 Rev. A
3152B User Manual
Base Line Offset
Adjustments
The base line offset adjustments assure that the AC signal is
symmetrical around the 0V line. Use this procedure if you suspect that
there is a base line accuracy issue.
Setup 12
Amplifier Offset, Amplifier In
Equipment:
DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
100
mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination
3. Configure the 3152B as follows:
Function Standard
Waveform
Sine
Amplitude
0
V
Frequency
1
MHz
Adjustment:
1. Adjust CAL:SETUP 2 for a DMM reading of 0 V
±
20 mV
Setup 13
Pre-Amplifier Offset, Amplifier Out
Equipment:
DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
100
mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination
3. Configure the 3152B as follows:
Function
Standard
Waveform
Sine
Amplitude
0
V
Frequency
1
MHz
Adjustment:
1. Adjust CAL:SETUP 3 for a DMM reading of 0 V
±
5 mV
Setup 14
Base Line Offset, Amplifier In – Modulation
Equipment:
DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
DCV
EADS North America Test and Services
Adjustments and Firmware Update 7-11