Publication No. 980935 Rev. A
3152B User Manual
Adjustment:
1. Adjust CAL:SETUP 54 for a DMM reading of 0 V,
±
100 mV
(Setup 6)
260 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
1
V
2. Configure the 3152B as follows:
SCLK:
260
MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 55 for a DMM reading of 0 V,
±
100 mV
(Setup 7)
277 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
1
V
2. Configure the 3152B as follows:
SCLK:
277
MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 56 for a DMM reading of 0 V,
±
100 mV
(Setup 8)
290 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
1
V
2. Configure the 3152B as follows:
SCLK:
290
MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 57 for a DMM reading of 0 V,
±
100 mV
EADS North America Test and Services
Adjustments and Firmware Update 7-7