Publication No. 980935 Rev. A
3152B User Manual
Amplitude
Adjustments
(Modulation)
The amplitude adjustments (modulation) ensure that AC levels are
within the specified range. Use this procedure if you suspect that the
amplitude accuracy is an issue for the CW modes.
Setup 42
10 V Amplitude, Amplifier In - Modulation
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
10
V
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination.
3. Configure the 3152B as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
10
V
Mode:
Modulation
Adjustment:
1. Adjust CAL:SETUP 26 for a DMM reading of 3.535 V
±
35 mV
Setup 43
5 V Amplitude, Amplifier In - Modulation
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
10
V
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination.
3. Configure the 3152B as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
5
V
Mode:
Modulation
Adjustment:
1. Adjust CAL:SETUP 27 for a DMM reading of 1.767 V
±
20 mV
EADS North America Test and Services
Adjustments and Firmware Update 7-23