5-36 Programming Reference
EADS North America Test and Services
would be 100e-3.
VOLTage:OFFSet <offs>(?)
HS Syntax: 0501A<float>(?)
Description
Programs the amplitude offset of the output waveform. The offset is calibrated when the source impedance is
50
Ω
.
Parameters
Name
Range
Type
Default
Description
<offs>
-7.992 to
7.992
Numeric
0
Sets the offset of the output waveform in units of Volts.
Offset and amplitude settings are independent providing
that the amplitude do not exceed the specified
window.
Response
The 3152B returns the current offset value. The returned value will be in scientific notation, e.g., 100 mV
would be 100e-3.
PHASe:OFFSet <phase_offs>(?)
HS Syntax: 0027A<float>(?)
Description
This command will affect a slave instrument only when it is synchronized to another module in the chassis. It
programs the start phase offset with respect to an adjacent master module. Phase offset resolution when
using this command is 1 point.
Parameters
Name
Range
Type
Default
Description
<phase_offs> 0 to 1e6-1
Numeric
(Integer
only)
0
Sets the phase offset in reference to a master
instrument. Slave instruments trail the master
instrument edge. The range is extended 4 Meg-1 when
the 4 Meg memory option is installed. 1 Meg is
standard.
Response
The 3152B returns the current phase offset value.
FUNCTion:MODE {FIXed|USER|SEQuence|SWEep|MODulation|
HALFcycle|COUNter}(?)
HS Syntax: 0610A<0|1|2|3|4|5|6>(?)
Description
This command selects the operating mode of the 3152B.
Parameters
Name
Type
Default
Description
FIXed
Discrete
FIX
Selects standard waveform mode. There is an
assortment of standard waveforms that may be
selected. You can find these waveform shapes in the
standard waveforms section.