Publication No. 980935 Rev. A
3152B User Manual
Frequency:
1
kHz
Output:
On
Amplitude:
100
mV
Offset:
1
V
Mode:
Modulation
Adjustment:
1. Adjust CAL:SETUP30 for a DMM reading of 35.35 mV
±
0.5 mV
Setup 47
50 mV Amplitude, Amplifier In - Modulation
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
100
mV
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination.
3. Configure the 3152B as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
50
mV
Offset:
1
V
Mode:
Modulation
Adjustment:
1. Adjust CAL:SETUP 31 for a DMM reading of 17.67 mV
±
2 mV
Setup 48
1 V Amplitude, Amplifier Out - Modulation
Equipment: DMM, BNC to BNC cable, 50
Ω
feedthrough termination,
Dual banana to BNC adapter
Preparation:
1. Configure the DMM as follows:
Function:
ACV
Range:
1
V
2. Connect the 3152B output to the DMM input. Terminate the
3152B output at the DMM input with the 50
Ω
feedthrough
termination.
3. Configure the 3152B as follows:
Frequency:
1
kHz
Output:
On
Amplitude:
1
V
Mode:
Modulation
Adjustment:
1. Adjust CAL:SETUP 32 for a DMM reading of 353.5 mV
±
3 mV
EADS North America Test and Services
Adjustments and Firmware Update 7-25