3152B User Manual
Publication No. 980935 Rev. A
Mod Run Mode: Gated
Carrier Freq:
1 MHz
Mod Frequency: 10 kHz
Deviation: 500
kHz
Sync: On
Output: On
Test Procedure:
1. Verify Gated FM – standard waveform operation on the
oscilloscope as follows:
Waveform:
Gated sine waveforms
Mod Frequency: 10 kHz
Gated Period: 1 ms
Max Deviation: 1.25 MHz
Min Deviation: 750 kHz
Test Results
Pass
Fail
Re-triggered FM
Bursts - Standard
Waveforms
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.2 ms
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2.
4. Configure model 3152B controls as follows:
Waveform:
Modulated
Modulation:
FM
Run Mode:
Burst
Burst Count:
5
Carrier Freq:
1 MHz
Mod Freq:
10 kHz
Deviation: 500
kHz
Sync: On
Re-trigger: On
Re-trig Delay: 200
μ
s
Output: On
Test Procedure:
1. Verify re-triggered FM burst – standard waveform operation
on the oscilloscope as follows:
Waveform:
Repetitive burst of 5-cycle sine waveforms
Mod Frequency: 10 kHz
Re-trigger delay: 200
μ
s
6-28 Performance Checks