Publication No. 980935 Rev. A
3152B User Manual
EADS North America Test and Services
Programming Reference 5-99
HALFcycle:FREQuency <freq>(?)
HS Syntax: 0140A<float>(?)
Description
Programs the frequency of the half cycle waveforms in units of hertz (Hz). It has no affect on the frequency of
other waveform functions.
Parameters
Name
Range
Type
Default
Description
<freq>
10e-3 to
1e6
Numeric
1e6
Sets the frequency of the half cycle waveform in units
of Hz. This parameter does not affect the frequency of
other waveform functions.
Response
The 3152B returns the current half cycle frequency value.
HALFcycle:PHASe <phase>(?)
HS Syntax: 0141A<float>(?)
Description
Programs the start phase of the half cycle sine and triangle waveform. This command has no affect on other
waveform functions.
Parameters
Name
Range
Type
Default
Description
<phase>
0 to 360
Numeric
0
Programs the start phase parameter for the half cycle
sine and triangle waveforms in units of degrees. The
phase can be programmable with resolution of 0.05
°
throughout the entire frequency range of the half cycle
function.
Response
The 3152B returns the current start phase value.
HALFcycle:SHAPe {SINusoid|TRIangle|SQUare}(?)
HS Syntax: 0142A<float>(?)
Description
This command defines the type of half cycle waveform that will be available at the output connector.
Parameters
Name
Type
Default
Description
SINusoid
Discrete
SIN
Selects the half cycle sine waveform.
TRIangle
Discrete
Selects the half cycle triangular waveform.
SQUare
Discrete
Selects the half cycle square waveform.
Response
The 3152B returns SIN, TRI, or SQU depending on the current 3152B setting
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com