3152B User Manual
Publication No. 980935 Rev. A
Frequency
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency:
As required by the test
Wave:
Square
Amplitude 500
mV
2. Connect the function generator to the 3152B TRIG IN connector.
3. Configure the 3152B, as follows:
Auxiliary Function: Counter/Timer
Function: Frequency
Trigger Level: 0 V
Test Procedure:
1. Perform Frequency Measurement Accuracy tests using
Table 6-28.
Table 6-28, Frequency Measurement Accuracy
Function
Generator Setting
Error Limits
3152B
Counter Reading
Pass
Fail
1.000000 MHz
±
2 Hz
100.0000 MHz
±
100 Hz
Period, Period
Averaged
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency:
As required by the test
Wave:
Square
Amplitude 500
mV
2. Connect the function generator to the 3152B TRIG IN connector
3. Configure the 3152B, as follows:
Auxiliary Function: Counter/Timer
Function: Period
Trigger Level: 0 V
Test Procedure:
1. Perform Period Accuracy tests using Table 6-29.
Table 6-29, Period Measurement Accuracy
Function
Generator Setting
Error Limits
3152B
Counter Reading
Pass
Fail
10 kHz
100.0
μ
s
±
100 ns
100 kHz
10.00
μ
s
±
100 ns
6-36 Performance Checks