3152B User Manual
Publication No. 980935 Rev. A
5-30 Programming Reference
EADS North America Test and Services
Parameters
Range
Type
Default
Description
0-1
Discrete
0
Toggles the SYNC output between on and off.
Response
The 3152B returns 1 if the SYNC output is on or 0 if the SYNC output is off.
OUTPut:SYNC:POSition<position>(?)
HS Syntax: 0242A<long>(?)
Description
Programs the 3152B SYNC position.
Parameters
Name
Range
Type
Default
Description
<position> 0 to 1e6-1
Numeric
(Integer only)
0
Sets the SYNC position in waveform points. The sync
position can be programmed in increments of 4 points
minimum (effectively 2 points in legacy mode). The
range is extended to 4e6-1 when the 4M memory option
is installed.
Response
The 3152B returns the current SYNC position value.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com