3152B User Manual
Publication No. 980935 Rev. A
Environmental
Conditions
The 3152B can operate from 0
°
C to 50
°
C. Adjustments should be
performed under laboratory conditions having an ambient temperature
of 25
°
C,
±
5
°
C and at relative humidity of less than 80%. Turn on the
power to the 3152B and allow it to warm up for at least 30 minutes
before beginning the adjustment procedure. If the instrument has been
subjected to conditions outside these ranges, allow at least one
additional hour for the instrument to stabilize before beginning the
adjustment procedure.
Warm-up Period
Most equipment is subject to a small amount of drift when it is first
turned on. To ensure accuracy, turn on the power to the Model 3152B
and allow it to warm-up for at least 30 minutes before beginning the
performance test procedure.
Recommended
Test Equipment
Recommended equipment for adjustments is listed in Table 7-1. Test
instruments other than those listed may be used only if their
specifications equal or exceed the required characteristics. Also listed
below are accessories required for calibration.
Table 7-1, Recommended Calibration for Adjustments
Equipment
Model No.
Manufacturer
Oscilloscope (with jitter package)
LC684D
LeCroy
Digital Multimeter
2000
Keithley
Frequency Counter (Rubidium reference) 2202R
Racal Instruments
Function Generator (with manual trigger)
WW2571A
Tabor Electronics
Accessories
BNC to BNC cables
“T”
BNC
adapter
Dual banana to BNC adapter
50
Ω
feedthrough termination
7-2 Adjustments and Firmware Update
EADS North America Test and Services
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com