Publication No. 980935 Rev. A
3152B User Manual
EADS North America Test and Services
Programming Reference 5-73
Modulation
Control
Commands
This group is used to control parameters for individual modulation
schemes. Control parameters are available for AM, FM, Sweep, FSK,
ASK, PSK, Amplitude and Frequency hopping, and 3D. The
modulation control commands are summarized in Table 5-10. Factory
defaults after *RST are shown in the Default column. Parameter range
and low and high limits are listed, where applicable.
Table 5-10, Modulated Waveform Control Commands
Keyword
Parameter Form
Default
3152A
HS
:AM
:FUNCtion
:SHAPe
SINusoid | TRIangle(*) | SQUare(*) | RAMP(*)
SIN (*) Computed
0057A
:INTernal
9
:FREQuency
10e-3
to
1e6
100
9
0781A
:DEPTh
0
to
100
50
9
0780A
:EXEcute
(=
MOD:TYPE
AM)
9
07820
:FM
:DEViation
10.0e-3
to
100e6
100e3
0075A
:FUNCtion
:SHAPe
SINusoid | TRIangle | SQUare | RAMP | ARB
SIN
0078A
:FREQuency
10e-3
to
350e3
10e3
0076A
:RASTer
1
to
2.5e6
1e6
0077A
:MARKer
[:FREQuency] 10e-3
to
100e6
1e6
0079A
:DATA
<data_array>
00741#
:SWEep
:FREQuency
[:STARt]
10
to
100.0e6
10e3
9
1610A
:STOP
10
to
100e6
1e6
9
1611A
:RASTer
10e-6 to 250e6 | MINimum | MAXimum
1e6
9
1612A
:FUNCtion
SINusoidal
|
TRIangle(*)
| SQUare(*)
SIN (*) Computed
9
1604A
:TIME
1.4e-6
to
40.0
1e-3
9
1600A
:DIRection
UP
|
DOWN
UP
9
1601A
:SPACing
LINear
|
LOGarithmic
LIN
9
1602A
:STEP
10
to
2000
20
9
1603A
:MARKer
[:FREQuency] 10
to
100e6
505e3
9
1613A
:FSK
:FREQuency
:SHIFted
10e-3
to
100e6
100e3
0082A
:BAUD
1
to
10e6
10e3
0080A
:MARKer
1
to
4000
1
0083A
:DATA
<data_array>
00811#