output start phase is shifted in reference to the external signal.
Parameters
Name
Range
Type
Default
Description
<phase>
-36 to 36
Numeric
0
Programs the phase offset between the reference and
the output waveform in smaller increments. The phase
is programmed in units of degrees (
°
). Note however,
that the phase offset resolution depends on the
number of points that create the waveform. For
example, waveform that is made of 1024 points can be
shifted with increments of 0.35
°
, but another waveform
that has only 100 points can be shifted with increments
of 3.6
°
.
Response
The 3152B returns the current fine phase offset value in units of degrees.
LXI System
Configuration
Commands
The LXI system configuration commands are available with the Model
3100M only (Message Based carrier) that has the LAN connector
installed on its front panel. Use these commands to configure module
address and other LAN parameters. It is strongly recommended that
this be done with a computer specialist because wrong programming
may place the instrument in an unknown configuration which may lock
the LAN operation completely and only hard reset will be necessary to
restore the instrument to its original defaults.
NOTE
Last LAN configuration settings will remain as long as
the instrument is turned on. New LXI configuration
settings will take effect only after the instrument has
been powered off and on.
Keyword
Parameter Form
Default
3152A
HS
:SYSTem
:IP
[:ADDRess]
<IP_address>
0167A
:MASK
<mask>
0171A
:GATeway
<gate_way>
0169A
:BOOTp
OFF | ON | 0 | 1
0
0168A
HOSTname:
<host_name>
0170A
:KEEPalive
:STATe
OFF
|
ON
|
0
|
1
1
0173A
:TIMEout
2
to
300
45
0174A
:PROBes
2
to
10
2
0172A
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com