3152B User Manual
Publication No. 980935 Rev. A
3-14
Operation
EADS North America Test and Services
sets the frequency to 10.7
MHz
(10.7e6), amplitude to 2 V, offset to
1 V, and duty cycle to 30%.
outp 1
Turns on the output.
The above is an example of a full utilization of the Apply command,
including the frequency, amplitude, offset, and duty cycle
parameters for a standard square wave. You may use the Apply
command in a similar manner for other standard or arbitrary
waveforms.
You may use the Apply command on a partial set of the available
parameters of a waveform. To leave out a parameter, place no
characters between the corresponding comma separators. For
example, you may specify the frequency and duty cycle of a square
wave while leaving the amplitude and offset at the default values,
as follows:
*rst
Restores factory defaults.
appl:squ 12.7e6,,,45
Selects the standard square wave as
the active function, and simultaneously
sets the frequency to 12.7 MHz, leaves
the amplitude at the default value,
leaves offset at the default value, and
sets the duty cycle to 45%.
outp 1
Turns on the output.
In a similar manner, you may use the Apply command with other
standard waveforms to set some parameters while leaving others at
their default values.
Generating
Standard
Waveforms
Each standard waveform is built into the 3152B in a lookup table or
equation. Ten standard function shapes are available:
•
Sine
•
Triangle
•
Square
•
Pulse
•
Ramp
•
Sinc
•
Exponential Decaying Pulse
•
Gaussian
Pulse
•
Noise
•
DC
Every time you select a standard function, the 3152B retrieves the
data points from a lookup table or calculates them from equations,
and then places them into waveform memory.
Use the following commands to select one of the standard