Publication No. 980935 Rev. A
3152B User Manual
Test Results
Pass
Fail
Fix Dwell Time
Frequency Hopping
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.1 ms
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2.
4. Configure model 3152B controls as follows:
Waveform:
Modulated
Modulation:
Hop
Hop Mode:
Fix
Dwell Time:
50
μ
s
Sync: On
Output: On
5. Using ArbConnection prepare, open the Hop Table composer
and download the following table:
Frequency
1.0e6
1.2e6
1.4e6
1.6e6
1.8e6
2.0e6
2.2e6
2.4e6
2.6e6
2.8e6
Test Procedure:
1. Verify Hop operation on the oscilloscope as follows:
Waveform:
Frequency steps, fixed dwell time of 50
μ
s
Max Hop Freq: 2.8 MHz
Min Hop Freq: 1.0 MHz
Period: 500
μ
s
Test Results
Pass
Fail
Amplitude Hopping
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
50
μ
s