Publication No. 980935 Rev. A
3152B User Manual
Max Deviation: 1.25 MHz
Min Deviation: 750 kHz
Test Results
Pass
Fail
FM - Arbitrary
Waveforms
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.2 ms
Sampling Rate: 50 MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
3. Connect the 3152B SYNC to the oscilloscope input, channel 2.
4. Configure model 3152B controls as follows:
Waveform:
Modulated
Modulation:
FM
Mod Waveform:Arbitrary
Carrier Freq:
1 MHz
FM SCLK:
2.5 MS/s
Sync: On
Output: On
5. Using ArbConnection, open the FM Composer and download
the following waveform:
Wavelength: 4000
points
Waveform:
4 sine wave cycles
Deviation: 0.5
MHz
Test Procedure:
1. Verify FM operation on the oscilloscope as follows:
Waveform: Sine
Mod Freq:
2.5 kHz
Max Deviation: 1.25 MHz
Min Deviation: 750 kHz
Test Results
Pass
Fail
AM
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.5 ms
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect the 3152B output to the oscilloscope input, channel 1.
EADS North America Test and Services
Performance Checks 6-29
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com