3152B User Manual
Publication No. 980935 Rev. A
Function:
DCV
Range:
1
V
2. Configure the 3152B as follows:
SCLK:
180
MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 51 for a DMM reading of 0 V,
±
100 mV
(Setup 3)
200 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
1
V
2. Configure the 3152B as follows:
SCLK:
200
MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 52 for a DMM reading of 0 V,
±
100 mV
(Setup 4)
220 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
1
V
2. Configure the 3152B as follows:
SCLK:
220
MHz
3. Connect the DMM probes between R107 and ground
Adjustment:
1. Adjust CAL:SETUP 53 for a DMM reading of 0 V,
±
100 mV
(Setup 5)
240 MHz SCLK
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Function:
DCV
Range:
1
V
2. Configure the 3152B as follows:
SCLK:
240
MHz
3. Connect the DMM probes between R107 and ground
7-6 Adjustments and Firmware Update