Publication No. 980935 Rev. A
3152B User Manual
SCLK 100
MS/s
Waveform: Sequence
Run Mode:
Trigger
Trigger Source: BUS
Amplitude: 2
V
Output: On
4. Using ArbConnection prepare and download the following
waveform:
Segments:
1 to 5
Wavelength: 128
points
Waveform:
1 cycle square
5. Using ArbConnection, build and download the following
sequence table:
Step 1:
Segment 1, loop 100,000
Step 2:
Segment 2, loop 100,000
Step 3:
Segment 3, loop 100,000
Step 4:
Segment 4, loop 100,000
Step 5:
Segment 5, loop 100,000
Test Procedure
1. From ArbConnection, click on the Manual Trigger button and
observe that counter reading is 500,000 counts. Reset
counter and repeat the test a few times. Every time the
counter reading should be 500,000 counts exactly.
Test Results
Pass
Fail
Step Advance
Equipment: Oscilloscope, function generator
Preparation:
1. Configure the Oscilloscope as follows:
Termination: 50
Ω
, 20 dB feedthrough attenuator at the
oscilloscope input
Setup:
As required for the test
2. Connect the 3152B output to the oscilloscope input.
3. Configure the function generator as follows:
Frequency 10
kHz
Run Mode:
Triggered
Waveform:
Square wave.
Amplitude:
Adjust for TTL level on 50
Ω
4. Connect the function generator output to the 3152B TRIG IN
connector.
5. Connect the 3152B to the Oscilloscope input.
6. Configure the 3152B as follows:
SCLK 100
MS/s
Waveform: Sequence
Seq. Advance: Step
Amplitude: 2
V
Trigger Source: External
Output: On