071207200
Instructions
AWG7000 SeriesArbitrary Waveform GeneratorsDeclassification and Security
071-2072-00
Страница 1: ...Revision A www tektronix com 071207200 071207200 Instructions AWG7000 Series Arbitrary Waveform Generators Declassification and Security 071 2072 00...
Страница 2: ...ished material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc Contacting Tektronix Tektronix Inc 14200 SW Karl Braun Drive or P O Box...
Страница 3: ...face iii Clear and Sanitize Procedures 1 Memory Devices 1 Media Storage and Data Export Devices 5 Built In Security Features 5 Troubleshooting 6 How to Clear or Sanitize a Non Functional Instrument 6...
Страница 4: ...Table of Contents ii AWG7000 Series Declassification and Security...
Страница 5: ...s also tell how to declassify an instrument that is not functioning Products The following Tektronix products are covered by this document AWG7101 and all options AWG7102 and all options AWG7051 and a...
Страница 6: ...clearing data putting the instrument in Standby mode does not qualify as powering off For these products you will need to either press a rear panel OFF switch or remove the power source from the inst...
Страница 7: ...odifiable1 Data input method Location Process to sanitize Removable hard drive Holds user storable data such as waveforms and instrument settings Yes Firmware operations Removable hard drive slot on r...
Страница 8: ...ollers entry by clicking the next to it 7 Double click the first USB Root Hub entry 8 Select the Power tab 9 If the Device Description is anything other than Generic USB Hub 4 ports click the General...
Страница 9: ...s Device Manager 1 Connect a PS2 mouse and a PS2 keyboard to the instrument before powering on because USB will be disabled 2 Log on to the instrument as an administrator 3 From the Windows Start menu...
Страница 10: ...e Tektronix has no recommendations regarding the available packages After the hard drive has been scrubbed the operating system and instrument software can be reinstalled using the Operating System Re...
Страница 11: ...ype Function User modifiable Data input method Location Process to disable USB host port Supports removable USB flash drive User storage of instrument setups and exported waveform data Yes User input...
Страница 12: ...rges Replacement of any missing hardware will be charged according to the rate at the time of replacement How to Recover from Clearing or Sanitizing the Instrument s Memory If you removed the boards i...