3.5 Operation screen reference
3
-
41
FRA 51615
■ [AMPLITUDE]
: Set the AC amplitude.
This is the amplitude of the signal that is output from the
OSC
connector. The unit is peak voltage (Vpk).
The setting range is 0 to 10 Vpk, but it is limited to ±10 V total, summed with the DC bias setting (
DC
BIAS
).
|DC bias setting (V)| + AC amplitude setting (Vpk)
≦
10.0 V
○ Setting method
< Numerical value settings (immediate setting) > , < Numerical value settings (new input) > ,
< Numerical value settings (edit) >
○ Function key functions
・
[DIRECT NEW]
: Set the amplitude with the numerical input dialog
【
NUM
】
.
・
[DIRECT EDIT]
: Edit the amplitude setting with the numerical input dialog
【
NUM
】
.
○ Functions of the function keys set with the numerical input dialog
【
NUM
】
・
[uVpk]
: Set with the entered value in units of µVpk.
・
[mVpk]
: Set with the entered value in units of mVpk.
・
[Vpk]
: Set with the entered value in units of Vpk.
(Additional explanation)
Even if the AC amplitude setting is changed, the output voltage does not change until the
AC/DC ON
key is pressed.
■ [DC BIAS]
: The DC bias setting
The DC bias (direct-current voltage) is set.
The setting range is –10 to +10 V, but it is limited to ±10 V together with the AC amplitude setting
(
AMPLITUDE
).
|DC bias setting (V)| + AC amplitude setting (Vpk)
≦
10.0 V
○ Setting method
< Numerical value settings (immediate setting) > , < Numerical value settings (new input) > ,
< Numerical value settings (edit) >
○ Function key functions
・
[DIRECT NEW]
: Set the DC bias with the numerical input dialog
【
NUM
】
.
・
[DIRECT EDIT]
: Edit the DC bias with the numerical input dialog
【
NUM
】
.
○ Functions of the function keys set with the numerical input dialog
【
NUM
】
・
[mV]
: Set with the entered value in units of mV.
・
[V]
: Set with the entered value in units of V.
(Additional explanation)
・
Even if the AC amplitude setting is changed, the output voltage is not changed unless the
AC/DC ON
key is pressed.
・
The DC bias is output to either the OSC connector on the front panel or to the
DC BIAS OUT
connector on the rear panel. The DC bias is output only to the selected connector. The output
connector is set by [
OUTPUT CONNECTOR
].
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......