9.4 Analysis processing section
9
-
15
FRA 51615
Measurement accuracy for other than Z and θ
Accuracy of measurement items other than Z (impedance) and θ (phase). Where defined the accuracy of
Z as Za %, and the measurement value of θ as θx, specified in "
9.4.1 Common items
, ● Measurement
accuracy (fixed range) and ● Measurement accuracy (AUTO range)".
Y
±Za %
R, Rs, Rp, G
±(Za / cos θx) %
X, B, Ls, Lp, Cs, Cp
±(Za / sin θx) %
D
±(0.01×Za)
When |Dx|
≦
0.1; Dx is the measured value of D.
The measurement accuracy of D is not a percent value, but the value itself.
Q
±
{(Qx
2
×
Pe / (1
-
|Qx|
×
Pe)}
When (|Qx|
≧
10) and (|Qx|
×
Pe
≦
0.1).
Qx is the measured value of Q. Pe = 0.01 × Za.
The measurement accuracy of Q is not a percent value, but the value itself.
Error correction function
Open correction
Corrects the residual admittance
Short correction
Corrects the residual impedance
Load correction
Corrects the voltage-current conversion coefficient of the measurement
system
Load standard value Standard values can be entered for up to 10 frequency points.
Port extension
Corrects the error due to phase delay in cables for 2-port measurements.
Characteristic impedance
Setting range
1.00 Ω to 999 Ω, resolution 3 digits
Electrical length
Setting range
0.000 m to 999.999 m, resolution 0.001 m
Slope compensation
ON/OFF
This function performs analysis that is unaffected by the DC level for signals
that have a superimposed DC level that varies linearly over time. It is used
when measuring the impedance of batteries during charging and discharging.
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......