5.9 Error correction
5
-
38
FRA 51615
5.9.4 Equalizing
This function corrects for errors of externally connected cables and probes, etc. This function is used in
gain-phase measurement.
The method of correcting for systematic error is explained with the connection example shown in
Figure
5-20
.
Figure 5-20
Example of connections for gain-phase measurement
The set up and measurement procedure is described below.
1) Set the measurement conditions, including the sweep frequency range, measurement range, and
oscillator amplitude, etc.
2) Set
【
Calibration
】
–
[EQUALIZING]
to
[OFF]
.
3) Remove the equipment under test (EUT) and connect the input and output.
Figure 5-21
Example of connections for systematic error
4) Turn the oscillator output on and perform a sweep measurement under the same conditions as when
measuring the EUT.
In
Out
EUT
OSC
CH1
CH2
FRA51615
Amp
Probe1
Probe2
BNC adapter
(provided)
OSC
CH1
CH2
FRA51615
Amp
Probe1
Probe2
BNC adapter
(provided)
Remove EUT
and connect
directly
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......