![NF FRA51615 Скачать руководство пользователя страница 266](http://html1.mh-extra.com/html/nf/fra51615/fra51615_instruction-manual_1684898266.webp)
9.1 Oscillator section
9
-
2
FRA 51615
Unless stated specifically otherwise, the assumed conditions are sinusoidal waveform, a temperature of
23°C ±5°C, no more than 12 hours has passed since calibration has been performed after a warm-up period
of at least 30 minutes, and change in ambient temperature of not greater than ±5°C after calibration.
Unless stated specifically otherwise, the shell (outer cover) and signal grounds are connected to the
instrument enclosure (not isolated).
9.1 Oscillator section
The oscillator section can output an AC signal and the DC bias for measurement. The AC signal is output
from the
OSC
connector on the front panel. The DC bias can be output from either the
OSC
connector on
the front panel or the
DC BIAS OUT
connector on the rear panel. You can select the connector from which
the DC bias is output. If the
OSC
connector on the front panel is selected for DC bias output, the signal
output from the
OSC
connector is a composite of the AC signal and the DC bias.
9.1.1 Front panel OSC
Connector
Insulated BNC connector (front panel,
OSC
)
Output waveform
Sinusoidal, square, or triangular
Frequency
Setting range
10 µHz to 15 MHz
The quality of square and triangular waveforms deteriorates at frequencies
above 100 kHz.
Setting resolution
10 µHz
Accuracy
±10 ppm
Conditions: The instrument is operating on the internal reference clock.
AC signal amplitude
Setting range
0 to 10 Vpk Conditions: No load
Setting resolution
3 digits or 0.01 mVpk, whichever is greater
Accuracy
±0.3 dB
Conditions: sinusoidal waveform, 1 kHz, 100 mVpk to 10 Vpk, no load
Frequency characteristic ±0.3 dB or less (100 kHz or less)
±1 dB or less (1 MHz or less)
±3 dB or less (15 MHz or less)
Conditions: 1 kHz reference, sinusoidal waveform, 100 mVpk to 10 Vpk, 50
Ω load
Distortion (sinusoidal) 0.2% or less, supplementary value
Conditions: 100 kHz or less, BW 500 kHz, 10 Vpk output
DC bias
The DC bias output is superimposed on the AC signal only when the DC bias output is set to the
OSC
connector on the front panel of the instrument.
Setting range
-10 V to 10 V
Setting resolution
10 mV
Accuracy
±(|1% of DC bias setting | + 2% of AC amplitude se 30 mV)
Conditions: sinusoidal waveform, no load
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......