4.2 Simple operation for the first use
4
-
12
FRA 51615
The voltage of the DUT current i × Rs is input to CH2 from the current detection resistance Rs. By setting
the CH2 input weight coefficient to 1/Rs and measuring the ratio of CH1 and CH2, the impedance value
can be read directly as the gain (ratio voltage and current). For more information on the input weight
coefficient, read "
5.9.2 Input weighting
".
Follow the procedure described below for instrument set-up and performing measurements.
1) Initialize settings.
2) Set the display unit to impedance.
Set
【
Graph
】
-
[DISPLAY UNIT]
to IMPEDANCE.
3) Set the graph Y axis display item.
Select with
【
Graph
】
dialog
-
[DISPLAY MODE][Y AXIS]
. For more information, read "
3.5.4 Graph
dialog
".
4) Press the
BASIC SETUP
key and change the settings as necessary.
[SWEEP LOWER]
: Sweep lower frequency
[SWEEP UPPER]
: Sweep upper frequency
[AMPLITUDE]
: Measurement amplitude
[INTEGRATION]
: Number of integrations
5) Press the
AC/DC ON
key on the front panel to turn the oscillator output on.
6) Press the
MEASURE DOWN
key on the front panel to perform a sweep measurement and wait until the
measurement is completed.
7) Turn the knob to move the marker on the measurement data. You can read the measured value at any
frequency. You can also detect maximum values, search, find the location of an arbitrary gain value, and
perform other such operations. For more information on the operations, read "
5.6 Marker operations
".
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......