3.5 Operation screen reference
3
-
57
FRA 51615
■ [DISPLAY MODE]
–
[Y AXIS]
: Set the formats of the Y1 and Y2 axes
This sets the formats for the Y1 and Y2 axes.
○ Setting method
< List operations (confirming) >
There are some formats that cannot be selected, depending on the data type of the Y axis. For more
information, refer to "
4.6 Graph display mode
".
When the
[DISPLAY UNIT]
is GAIN
• R – θ
:
The Y1 axis is gain (unitless number ), the Y2 axis is phase (°).
• dBR – θ
: The Y1 axis is gain (dB ), the Y2 axis is phase (°).
• R – GD
: The Y1 axis is gain (unitless number ), the Y2 axis is group delay (sec).
• dBR – GD
: The Y1 axis is gain (dB ), the Y2 axis is group delay (sec).
• a – b
: The Y1 axis is the real part of the gain (unitless number), the Y2 axis is the
imaginary part of the gain (unitless number).
• R
: The Y1 axis is gain (unitless number), the Y2 axis is nothing.
• dBR
: The Y1 axis is gain (dB), the Y2 axis is nothing.
• b
: The Y1 axis is the imaginary part of the gain (unitless number), the Y2 axis is
nothing.
When the
[DISPLAY UNIT]
is IMPEDANCE
• Z
-
θ
Z
: The Y1 axis is impedance (Ω), the Y2 axis is phase (°) of impedance.
• Y
-
θ
Y
: The Y1 axis is admittance(S), the Y2 axis is phase (°) of admittance.
• R
-
X
: The Y1 axis is resistance (Ω), the Y2 axis is reactance (Ω).
• G
-
B
: The Y1 axis is conductance(S), the Y2 axis is susceptance (S).
• C
p
-
R
p
: The Y1 axis is parallel capacitance (F), the Y2 axis is parallel resistance (Ω).
• C
p
-
D
: The Y1 axis is parallel capacitance (F), the Y2 axis is dissipation factor (unitless
number).
• C
p
-
Q
c
: The Y1 axis is parallel capacitance (F), the Y2 axis is quality factor (unitless
number).
• C
s
-
R
s
: The Y1 axis is series capacitance (F), the Y2 axis is series resistance (Ω).
• C
s
-
D
: The Y1 axis is series capacitance (F), the Y2 axis is dissipation factor (unitless
number).
• C
s
-
Q
c
: The Y1 axis is series capacitance (F), the Y2 axis is quality factor (unitless
number).
• L
p
-
R
p
: The Y1 axis is parallel inductance (H), the Y2 axis is parallel resistance (Ω).
• L
p
-
Q
L
: The Y1 axis is parallel inductance (H), the Y2 axis is quality factor (unitless
number).
• L
s
-
R
s
: The Y1 axis is series inductance (H), the Y2 axis is series resistance (Ω).
• V
: The Y1 axis is voltage (Vrms) , the Y2 axis is nothing.
• I
: The Y1 axis is current (Arms) , the Y2 axis is nothing.
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......