3.5 Operation screen reference
3
-
72
FRA 51615
■ [WEIGHTING FACTOR]
–
[CH1]
: Set the CH1 weight factor.
■ [WEIGHTING FACTOR]
–
[CH2]
: Set the CH2 weight factor.
This value serves as a multiplier for the measurement result. It can be used as a simple correction factor
for when a probe, amplifier, or shunt resistance is connected to the circuit being measured.
○ Setting method
< Numerical value settings (immediate setting) > , < Numerical value settings (new input) > ,
< Numerical value settings (edit) >
○ Function key functions
• [DIRECT NEW]
: Set the maximum or minimum X axis value with the numerical input
dialog
【
NUM
】
.
• [DIRECT EDIT]
: Edit the maximum or minimum X axis value with the numerical input
dialog
【
NUM
】
.
○ Functions of the function keys set with the numerical input dialog
【
NUM
】
• [u(10
– 6
)]
: Set with the entered value in units of 10
– 6
.
• [m(10
– 3
)]
: Set with the entered value in units of 10
– 3
.
• [×1]
: Set with the entered value.
• [k(10
3
)]
: Set with the entered value in units of 10
3
.
• [M(10
6
)]
: Set with the entered value in units of 10
6
.
• [p(10
– 12
)]
: Set with the entered value in units of 10
– 12
.
• [n(10
– 9
)]
: Set with the entered value in units of 10
– 9
.
• [G(10
9
)]
: Set with the entered value in units of 10
9
.
• [T(10
12
)]
: Set with the entered value in units of 10
12
.
The setting range is 0.0 to 1.0 T (10
12
) and the resolution is 6 digits or 1 p(10
– 12
), whichever is larger.
■ [WEIGHTING FACTOR]
–
[INVERT]
: Set phase inversion.
This inverts the phase of the measurement results.
○ Setting method
< List operations (confirming) >
• ON
: Display the measurement results with phase inversion.
• OFF
: Do not invert phase.
This is set to ON when measuring imepedance with the optional
"PA-001-0368 IMPEDANCE
MEASUREMENT ADAPTER".
For more information, refer to "
5.9.3 Input weighting
".
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......