8.5 Performance test
8-9
FRA 51615
8.5.5 Oscillator DC bias accuracy (OUTPUT OSC)
This instrument
: Set the AC amplitude to 0 Vpk.
Measuring instrument
: Set the digital multimeter to DC voltage mode.
Connections
: Connect the
OUTPUT OSC
connector of this instrument to the digital
multimeter with a coaxial cable.
Measurement method
: Set the DC bias of this instrument to −10 V, 0 V, and +10 V, and read the DC
voltage from the multimeter.
Judgement
DC bias setting
DC voltage value
Passing range
−10 V
− _ _. _ _ _ V
−10.130 to −9.870 V
0 V
_ _ _. _ mV
−30.0 to +30.0 mV
+10 V
+ _ _. _ _ _ V
+9.870 to +10.130 V
FRA 51615
OSC
CH1
CH2
Digital multimeter
BNC-BNC
coaxial cable
BNC-banana
adapter
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......