![NF FRA51615 Скачать руководство пользователя страница 209](http://html1.mh-extra.com/html/nf/fra51615/fra51615_instruction-manual_1684898209.webp)
5.9 Error correction
5
-
47
FRA 51615
Measure the open correction data. Refer to "
5.9.6 Open correction
".
3) Open the measurement port circuit. (Refer to
(a)
in "
Figure 5-24 Connections for open
correction and short correction
".) Keep the spacing and positions at the measurement ports the
same as for when performing DUT measurement.
4) Turn the oscillator output on and perform a sweep measurement. Wait until the measurement is
completed.
5) Select
【
Trace
】
-
[MEAS COPY]
and tap on the
[OPEN CORR MEM]
function key to copy the
correction measurement data to the open correction memory.
Measure the short correction data. Refer to "
5.9.7 Short correction
".
6) Short the measurement ports (Refer to
(b)
in "
Figure 5-24 Connections for open correction and
short correction
".) Keep the spacing and positions at the measurement ports the same as for when
performing DUT measurement and short the ports with a metal plate that has low inductance.
7) Turn the oscillator output on and perform a sweep measurement. Wait until the measurement is
completed.
8) Select
【
Trace
】
-
[MEAS COPY]
and tap on the
[SHORT CORR MEM]
function key to copy the
correction measurement data to the short correction memory.
Measure the load correction data and input the load standard value.
9) Connect the impedance standard device to the measurement ports. Keep the spacing and positions
at the measurement ports the same as for when performing DUT measurement.
10) Turn the oscillator output on and perform a sweep measurement. Wait until the measurement is
completed.
11) Select
【
Trace
】
-
[MEAS COPY]
and tap on the
[LOAD CORR MEM]
function key to copy the
correction measurement data to the load correction memory.
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......