2.4 Quick operation check
2
-
13
FRA 51615
2.4 Quick operation check
If the instrument has not been used for a long time after purchase, you should check instrument operation
by the procedure described below. Prepare the two calibration cables and the BNC adapter that are supplied
with the instrument.
For a description of the labels and operation of the keys and screen displays, etc., please refer to "
3.1
Function and use of each part of the panels
" and "
3.3 Operation screens
".
1. After connecting the instrument to the power supply, press the standby switch on the front panel and then
wait until the system check ends normally.
About the appearance of the display after the power has been turned on
→
"
4.1 Display and operation for powering up
".
2. Initialize the settings.
Press the
MENU
key two times and then tap on [
Other
] on the second page of the top menu. Select
【
Other
】
- [
RESET
] and then press the
ENTER
key. When the confirmation message is displayed, tap
on the [
YES
] button.
3. Plug the provided BNC adapter into the
OUTPUT
OSC
connector. Plug the provided calibration cables
into the BNC adapter and connect them to
INPUT
CH1
and
INPUT
CH2
as shown in the following
diagram.
4. Run the calibration (self-measurement for error correction).
Tap on [
Calib.
] on the second page of the top menu. Select
【
Calibration
】
- [
CALIBRATION
] and then
press the
ENTER
key. When the cable connection confirmation message is displayed, tap on the [
OK
]
button and then wait until the calibration is completed.
For information on calibration
→
"
5.9.1 Calibration
".
5. Turn the oscillator output on.
On the front panel, press the ON
AC/DC
key to turn on the oscillator output.
The AC lamp and DC lamp to the right of the ON
AC/DC
key both light up.
6. Perform a sweep measurement to confirm that correct measurement is possible.
On the front panel, press the
MEASURE START
UP
key to perform an up-sweep measurement.
When the measurement is finished, a graph display of the gain and phase is presented. The normal values
are 0 dB ±0.05 dB for the gain and 0 deg ±3.0 deg for the phase.
This concludes the simple operation check.
OUTPUT
OSC
INPUT
CH1
CH2
BNC adapter
Calibration cables
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......