9.4 Analysis processing section
9
-
11
FRA 51615
Inter-range accuracy
Upper: gain (ratio); Middle: impedanse Z; Lower: phase
Measurement range
(rms)
Frequency
≦
100 kHz
≦
200 kHz
≦
2 MHz
≦
15 MHz
600 V
-
±0.1 dB
±1.2 %
±0.6°
-
-
-
300 V
±0.05 dB
±0.58 %
±0.3°
±0.1 dB
±1.2 %
±0.6°
-
-
100 V
±0.05 dB
±0.58 %
±0.3°
-
-
30 V
±0.05 dB
±0.58 %
±0.3°
-
10 V
±0.05 dB
±0.58 %
±0.3°
3 V
±0.05 dB
±0.58 %
±0.3°
1 V
±0.05 dB
±0.58 %
±0.3°
300 mV
±0.05 dB
±0.58 %
±0.3°
100 mV
-
Conditions: At least 30 cycles of integration and fixed measurement range for both channels.
The gain and phase error for when the measurement range difference between channels
is 1, the input signal levels of both channels are equal, and equal to the range full scale
level of the smaller range.
*For the cells that contain only " – ", the measurement range exceeds the maximum input voltage (see
"
9.2 Analysis input section
, ● Maximum input voltage"), so either measurement is not possible or
there is no accuracy specification for it.
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......