4.2 Simple operation for the first use
4
-
13
FRA 51615
If the OSC output of the instrument is amplified with a power amplifier, it is possible to measure the
impedance by applying a large voltage/current to the DUT. A large amplitude or DC bias (voltage/current)
can be used to measure a change in the impedance characteristic caused by a nonlinearity such as saturation
of a magnetic or dielectric material.
Figure 4-7
Example of connections for measuring impedance with high-power
The wiring and connections are simplified by using the optional "
PA-001-1840 HI-POWER
IMPEDANCE MEASUREMENT ADAPTER (1 ohm)"
or
the
"PA-001-1841 HI-POWER
IMPEDANCE MEASUREMENT ADAPTER (100 ohm)".
For more information, refer to the instruction
manual of each adapter.
1
~
2
Zx
Current detection
resister
Rs
i
i x Rs
v
Power amplifier
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......