4.2 Simple operation for the first use
4
-
11
FRA 51615
4) Press the
MEASURE DOWN
key on the front panel to perform a sweep measurement and wait until the
measurement is completed.
5) Turning the knob moves the marker on the measurement data. You can read the gain or phase at any
frequency. You can also detect maximum values, search, find the location of an arbitrary gain value, and
perform other such operations. For more information on the operations, read "
5.6 Marker operations
".
4.2.4 Measuring impedance
Impedance is defined as the ratio of the voltage and current of the device under test (DUT). The current of
the sample is converted to voltage using a shunt resister. By measuring the voltage of the shunt resister and
the sample simultaneouly, the impedance of the sample can be obtained. The DUT voltage and the voltage
converted from the DUT current are connected to CH1 and CH2 respectively..
Figure 4-6
Example of connections for measuring impedance
This instrument can be easily connected to DUTs by using the optional "
PA-001-0368 IMPEDANCE
MEASUREMENT ADAPTER"
. For more information, please refer to the instruction manual of
"
PA-001-0368 IMPEDANCE MEASUREMENT ADAPTER
".
1
~
2
Current detection
resister
Rs
i
i x Rs
v
DUT
Zx
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......