8.5 Performance test
8-5
FRA 51615
3) Testing equipment
The measurement instruments needed for checking operation and testing performance are listed as below.
Product
Required specifications
Example models
Frequency counter
Accuracy: ±1 ppm or better
53131A
+ option 001
(Keysight)
Digital multimeter
AC voltage (true rms)
(100 mV to 10 V, 100 Hz to 10 kHz)
Accuracy: ±0.1%
DC voltage
Accuracy: ±0.1%
(100 mV to 10 V)
34401A
(Keysight)
RF voltmeter
AC voltage (true rms)
(100 mV to 10 V)
Accuracy: ±0.5%
(1 kHz to 100 kHz)
Accuracy: ±1.0%
(100 kHz to 1 MHz)
Accuracy: ±5.0%
(1 MHz to 15 MHz)
URE3
(Rohde & Schwarz)
Other
BNC-BNC coaxial cable 50 Ω, RG-58A/U
× 2
BNC 50 Ω feed-thru terminator
× 1
(Power rating is 0.25 W or more)
*1
BNC-banana adapter
× 1
BNC adapter (provided)
× 1
BNC short plug
× 1
BNC-alligator clip cable
× 1
*1: Not required if the input impedance of the RF voltmeter is 50 Ω.
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......