5.4 Oscillator control
5
-
11
FRA 51615
5.4.4 Measurement synchronization
This function automatically turns the oscillator output off when the measurement is completed. The setting
is done with
【
OSC
】
–
[ON/OFF SYNC]
. This function is used when it is not desirable for the measurement
signal to be applied to the equipment under test for a long period of time.
[ASYNC]
: The oscillator output is not turned off, even when the measurement is completed.
[SYNC]
: The oscillator output is turned off when the measurement is completed.
The order of operations for when the setting is SYNC is shown below.
Start measurement (Press the
UP
,
DOWN
, or
SPOT
key.)
↓
Oscillator output is turned on.
↓
Wait for the START DELAY time. (Measurement is not performed during this time.)
↓
Start signal acquisition (begin analysis)
↓
(frequency changes during the sweep)
:
End analysis (the sweep ends)
↓
Oscillator output is turned off
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......