4.3 Oscillator settings
4
-
15
FRA 51615
The frequency is set by tapping the function keys at the bottom of the LCD screen.
[SWEEP MIN -> FREQ]
: Set the lower limit for the sweep frequency.
[SWEEP MAX -> FREQ]
: Set the upper limit for the sweep frequency.
[MKR -> FREQ]
: Set the main marker frequency.
[
⊿
MKR -> FREQ]
: Set the delta marker frequency.
[DIRECT NEW]
: Set an arbitrary frequency.
[DIRECT EDIT]
: Edit the current frequency setting.
Changing the frequency setting immediately changes the oscillator output frequency.
When a sweep measurement is performed, the setting changes according to the sweep frequency. When a
sweep is completed, the setting value becomes the frequency at the end of the sweep.
■ Setting the AC amplitude
The AC amplitude for measurement is set in the range from 0 to 10 Vpk.
Select
【
OSC
】-
[AMPLITUDE]
.
The AC amplitude is set by tapping the function keys at the bottom of the LCD screen.
[DIRECT NEW]
: Set an arbitrary frequency.
[DIRECT EDIT]
: Edit the current frequency setting.
● Cautions concerning the AC amplitude setting
Simply setting a value for the AC amplitude does not change the actual amplitude of the oscillator
output until the
AC/DC ON
key is pressed. Even if the AC and DC output lamps are lit, it is necessary
to press the
AC/DC ON
key to change the actual output amplitude.
● Constraints on the AC amplitude setting range and the DC bias setting range
The settings for the AC amplitude and the DC bias are limited to a total of ±10 V. If the DC bias is set
to 0 V, the AC amplitude can be set to a value of up to 10 Vpk. Please note that if the DC bias is set to
+1.2 V (or −1.2 V), the AC amplitude setting range is restricted to a maximum of 8.8 Vpk.
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......