9.2 Analysis input section
9
-
5
FRA 51615
At 45 Hz and above, the voltage including DC must be 600 Vrms or less and
849 Vpk or less.
Capacitance relative to enclosure
200 pF or less
Frequency range
10 µHz to 15 MHz
Maximum measurement voltage
600 Vrms
When the provided signal cable is used.
When the other signal cables are used, the voltage is restricted to the rated
voltage of the cable or to 600 Vrms, whichever is less.
Over-level detection
Setting range
0 to 600 Vrms
Setting resolution
3 digits
Processing
Over lamp lights.
Buzzer warning sound (can be turned on / off)
Stop sweep measurement (can be turned enable / disable)
Over-level detection occurs when the amplitude or maximum measurement input voltage that is set as
described above is exceeded. (See "
9.2 Analysis input section
, ● Measurement range”.)
Dynamic range
140 dB (10 Hz to 1 MHz)
80 dB (1 MHz to 15 MHz)
The maximum measurable gain, limited by the noise floor and cross talk
between channels.
Conditions: The larger channel input is at least 10 Vpk, integration is 300
cycles, and the measurement range is AUTO for both channels.
Input weighting
Setting range is 0 to 1.12, resolution 6 digits or 1E-12
CH1 and CH2 are set independently
The weight is a coefficient that is applied to measurement results of CH1 or
CH2 to correct the attenuation or gain of probes, amplifiers, or other such
equipment that is connected to the instrument.
Phase inversion function ON or OFF
This function is used when the polarity of connection to the circuit being
tested is reversed. This is set to on when the optional "
PA-001-0368
impedance measurement adapter
" is used.
The isolation specifications for between the oscillator (
OSC
) and analysis input (
CH1
or
CH2
) and the
instrument enclosure is illustrated in Figure 9-1. The isolation specifications between the
OSC
,
CH1
, and
CH2
signal connectors is illustrated in Figure 9-2.
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......