9.4 Analysis processing section
9
-
9
FRA 51615
9.4 Analysis processing section
The SI prefixes (k, M, etc.) that are placed before units of measurement and the scales they represent are a
(10
-18
), f (10
-15
), p (10
-12
), n (10
-9
), µ (10
-6
), m (10
-3
), k (10
3
), M (10
6
), and G (10
9
).
9.4.1 Common items
Display unit
Gain (ratio, unitless number) or impedance
Measurement accuracy (fixed range)
Measurement accuracy Relative ac Calibration accuracy
Relative accuracy
± (| Basic accuracy | + | Dynamic accuracy | + | Inter-range accuracy × N|),
where N is | number of measurement range steps between channels |.
Example: If the measurement ranges of each channels are 300 Vrms and 1
Vrms, N is 5 (300 V to 100 V to 30 V to 10 V to 3 V to 1 V).
Calibration accuracy
The accuracy of external equipment that is connected to the instrument, such
as a shunt resistor or probe, or the accuracy of the calibration standard
equipment.
Basic accuracy
Upper: gain (ratio); Middle: impedance Z; Lower: phase
Measurement range
(rms)
Frequency
≦
100 kHz
≦
200 kHz
≦
1 MHz
≦
2 MHz
600 V
±0.2 dB
±2.4 %
±1.2°
-
-
-
300 V
±0.1 dB
±1.2 %
±0.6°
-
-
100 V
±0.05 dB
±0.58 %
±0.3°
-
-
30 V
:
30 mV
±0.01 dB
±0.12 %
±0.06°
±0.025 dB
±0.29 %
±0.15°
±0.1 dB
±1.2 %
±0.6°
Measurement range
(rms)
Frequency
≦
5 MHz
≦
15 MHz
10 V
:
30 mV
±0.2 dB
±2.4 %
±1.2°
±0.5 dB
±5.9 %
±3.0°
Conditions: At least 30 cycles of integration, fixed measurement range for both channels, and
both channels have the same range. The gain and phase error for when the signal
input is at the full scale of the measurement range for both channels.
*For the cells that contain only " – ", the measurement range exceeds the maximum input voltage (see
"
9.2 Analysis input section
, ● Maximum input voltage"), so either the measurement is not possible
or there is no accuracy specification for it.
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......