6.2 Measurement data file format
6
-
5
FRA 51615
30m}
+ 21
InputRangeCh2, {AUTO|600|300|
100|30|10|3|1|300m|100m|
30m}
The CH2 measurement range
The line number of the first line of the measurement conditions part relative to the beginning of the file can
be calculated with the formula shown below, denoting the number of data items (DataNum) for sequence
number m as n
m
.
First line number for the measurement conditions part = 4 + (21 + n
1
) + (21 + n
2
) + … + (21 + n
m
)
If there is no sequence measurement (m = 0), the line number of the first line of the measurement
conditions part would be 4.
Table 6-3
Data file formats (measurement data part)
Relative
line number
Format (example)
Description
+ 0
(NR2), (NR3), (NR3),
(NR3), (NR3), (NR3)
RAW[0]
Frequency (Hz), V1 (Vrms), V2 (Vrms),
θ (deg), Y1, Y2
+ 1
(NR2), (NR3), (NR3),
(NR3), (NR3), (NR3)
RAW[1]
Frequency (Hz), V1 (Vrms), V2 (Vrms),
θ (deg), Y1, Y2
:
:
:
+ (n-1)
(NR2), (NR3), (NR3),
(NR3), (NR3), (NR3)
RAW[n-1]
Frequency (Hz), V1 (Vrms), V2 (Vrms),
θ (deg), Y1, Y2
○ Concerning Y1 and Y2
These refer to the Y1 axis and Y2 axis of the graph. According to the graph settings at the time the file was
saved, the contents are stored in the relative line number 19 (Graph GraphYaxis on the relative line number
19 on the measurement conditions part. Unless the graph X axis is SWEEP, Y1 is the X axis and Y2 is the
Y axis.
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......