3.5 Operation screen reference
3
-
43
FRA 51615
■ [STOP MODE]
: Set the oscillation stop phase.
This selects whether to stop the oscillation instantaneously or to wait until the 0° phase of a sine function
before stopping the oscillation. This is useful for the measurement of electrochemical impedance because
the amount of charge transfer can be made zero by equalizing the areas of the positive and negative sides
of an AC sinusoidal signal.
○ Setting method
< List operations (confirming) >
• QUICK
: Turn off instantly.
• 0° SYNC
: Turn off after waiting for 0° phase (stop oscillation).
When
0° SYNC
is set, it takes up to one cycle of the signal at the longest from the time the output off
operation is performed to the time the output is actually turned off. Please note that, at 10 µHz, it takes up
to 10
5
seconds (27.8 hours) at the longest.
If the
[ON/OFF MODE]
is set to
SLOW
and the
[STOP MODE]
is set to
0° SYNC
, the
[STOP
MODE]
setting has priority.
■ [WAVEFORM]
: Select the waveform.
This selects the oscillator output waveform from among sinusoidal, square, and triangular. Normally,
sinusoidal is used. Main characteristics such as amplitude accuracy and measurement accuracy are
specified only for the sinusoidal waveform.
○ Setting method
< List operations (confirming) >
• SINE
:
Sinusoidal
• SQR
: Square
• TRI
: Triangular
The square and triangular waveforms should be used for frequencies of 100 kHz and below. At higher in
frequencies, waveform quality decreases.
Содержание FRA51615
Страница 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS nbn Austria GmbH ...
Страница 2: ......
Страница 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Страница 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Страница 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Страница 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Страница 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Страница 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Страница 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Страница 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Страница 300: ...XVI FRA51615 Blank ...
Страница 302: ......
Страница 304: ......
Страница 305: ......