AquaMax
®
2000/4000 Microplate Washer User Guide
98
0112-0151 K
External Communication Issues
Table 6-4:
External Communication Issues
Problem
Possible Causes & Remedies
The StakMax
®
stacker is not
communicating.
Verify that the StakMax cable is connected to
both the StakMax stacker and the AquaMax
®
Microplate Washer.
Power-cycle both instruments to re-establish
the connections.
The robot is not
communicating.
Verify that the USB cable is connected to both
the robot and the AquaMax Microplate
Washer.
Verify that the communication port setting
corresponds to the virtual COM port assigned
to the USB driver.
Power-cycle both the robot and the AquaMax
Microplate Washer to re-establish
connections.
Data cannot be downloaded
from instrument to computer
using the AquaMax Utility.
Verify that the USB cable is plugged into both
the AquaMax Microplate Washer and the
computer.
Verify that the communication port setting
corresponds to the virtual COM port assigned
to the USB driver.
Restart the computer and power-cycle the
AquaMax Microplate Washer to re-establish
connections.
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...